Analysis Setup

Calibration Programmes

A Calibration Programme declares, for a class of instruments, how often they are calibrated, how often they are checked in between, and when the programme itself is reconsidered (ISO/IEC 17025 §6.4.7, §6.4.10). It carries the §6.4.7 obligation that the programme be "reviewed and adjusted as necessary".

Found under: Analysis Configuration → Equipment → Calibration Programmes.

Toolbar

ButtonAction
Add — Create a new record
Columns — Show/hide table columns
Filters — Open/close the filter panel. A red dot indicates active filters
Export CSV — Download table data as a CSV file
Orientation — Toggle between portrait and landscape printing
Print — Generate a PDF of the current view
Help — Open the help dialog
Search — Quick search through table data

Record Actions

Clicking the icon on each row shows:

  • Edit — Open the edit form
  • Delete — Delete the record (with confirmation)

Fields

FieldDescription
NameHow the programme is referred to — required
Instrument ClassWhich instruments it governs; only active classes are listed
Interval (months)The default recalibration interval
Internal Check Required · Internal Check FrequencyWhether and how often intermediate checks are done (§6.4.10)
Next ReviewWhen the programme is reconsidered
JustificationWhy this interval holds — required when the interval is lengthened
Review NotesWhat the last review concluded
ActiveWhether the programme applies

When the interval gets longer

Lengthening a recalibration interval is not a simple change of number. ESYD KO1-KRITE §4.5 asks for two things together, and the system enforces both:

  1. A new justification. The text that justified the shorter interval is not enough — you must write a fresh one.
  2. More frequent intermediate checks. The internal check frequency must become strictly greater than it was.

If either is missing, the save is refused with a message saying which. Shortening an interval is always allowed without conditions: it is itself a corrective action.

The frequency order, from most to least frequent, is: Before Use, Per Shift, Daily, Per Batch, Weekly, Monthly, Quarterly, Semiannually, Annually.

Connects to

Points at — this page does not stand without these:

Used by — a change here shows up in these:

If this is missing

  • The programme is the central rule: instruments that point at no programme inherit no interval,

so the interval has to be typed into each one by hand. Any that is forgotten ends up with no next

calibration date — and then the overdue check never locks that instrument out (see

Instruments).

  • With no justification for the interval, ESYD's ΚΟ1-ΚΡΙΤΕ §4 is not satisfied: the assessor

asks why twelve months and not six. "That is how we found it" is not an answer.

  • Extending an interval with no recorded justification is the finding that gets written up.

Notes

  • The programme supplies the default. A particular instrument's next calibration date is computed on the instrument's own page and may use its own interval.
  • The date filters are labelled Created From/To but filter on Next Review.
  • See also: Instrument Classes, Instrument Validations, Correction Factors.