When a calibration shows that an instrument reads systematically wrong — "this thermometer reads 0.3 °C high" — the laboratory must record the correction and apply it to subsequent measurements (ISO/IEC 17025 §6.4.11). This page holds those factors and their history.
Found under: Analysis Configuration → Equipment → Correction Factors.
| Button | Action |
|---|---|
| Add — Create a new record | |
| Columns — Show/hide table columns | |
| Filters — Open/close the filter panel. A red dot indicates active filters | |
| Export CSV — Download table data as a CSV file | |
| Orientation — Toggle between portrait and landscape printing | |
| Print — Generate a PDF of the current view | |
| Help — Open the help dialog | |
| Search — Quick search through table data |
Clicking the icon on each row shows:
| Field | Description |
|---|---|
| Instrument | Which instrument is corrected — required |
| Element | Which element it applies to; blank means every element on the instrument |
| Type | How it is applied — see below |
| Value | The size of the correction — required |
| Unit | The unit of the value, for example °C, %, mg/L |
| Effective From | When it starts to apply — required |
| Effective To | Filled in automatically when a later factor supersedes it |
| Notes | Additional remarks |
| Type | How it is applied |
|---|---|
| Additive | measurement + value |
| Offset | measurement − value |
| Multiplicative | measurement × value |
Correction factors are never edited destructively. Each time you record a factor for an instrument and element pair, the previous ones close automatically: they take an end date of the day before and point at the new one. An old measurement can therefore always be re-justified with the factor that was in force at the time.
Supersession respects the element: a factor for a specific element supersedes only others for that element, and a factor with no element supersedes only others with no element.