Lab Methods

The Lab Methods page manages the laboratory's analytical methods: the detection limits, the SOP, the metrological traceability, and how often the method is checked.

Found under: Analysis Configuration → Methods & Analyses → Methods.

Toolbar

ButtonAction
Add — Create a new record
Columns — Show/hide table columns
Filters — Open/close the filter panel. A red dot indicates active filters
Export CSV — Download table data as a CSV file
Orientation — Toggle between portrait and landscape printing
Print — Generate a PDF of the current view
Help — Open the help dialog
Search — Quick search through table data

Record Actions

Clicking the icon on each row shows:

  • Edit — Open the edit form
  • Delete — Delete the record (with confirmation)

Fields

FieldDescription
MethodMethod name (for example Olsen, DTPA, Walkley-Black)
Description GRMethod description in Greek
Description ENMethod description in English
LODLimit of Detection (hidden column)
LOQLimit of Quantitation (hidden column)
UncertaintyExpanded measurement uncertainty (hidden column)
SOP VersionCurrent SOP version number (hidden column)
SOP Approval DateDate the SOP was approved (hidden column)
SOP ContentThe text of the procedure
Traceability Pathway (§6.5.3)What the results are traceable to — see below
QC interval (days)§7.7.1 — how many days may pass between internal quality checks
PT interval (months)§7.7.2 — how many months may pass between interlaboratory comparisons
Next Method Review (§7.2.1.6)When the method must be looked at again; the last review shows below the field
AccreditationWhether the method is accredited (Yes/No)

Traceability pathway (§6.5.3)

Every method must state what its results are traceable to. The pathway is one of: the SI, a certified reference material, a reference method, or a consensus standard. The field records which one; the rationale for the choice is documented in the SOP.

The two check intervals

The two intervals are not calendar reminders: they are the two silent arms of §7.7. A method that has not run an internal quality check within its QC interval, or has not taken part in an interlaboratory comparison within its PT interval, raises no other signal by itself — which is why both feed the quality-control alerts.

Connects to

Points at — this page does not stand without these:

Used by — a change here shows up in these:

Notes