A lab has two completely different kinds of limits that look alike on paper (both are just
numbers — a lower and an upper limit), but they **answer a different question about a different
thing**. Confusing them is the single most common source of mistakes — which is why this LIMS
keeps them on two separate screens.
A lab scale:
scale is fine. → "Is my method working correctly?" That's a Control-Chart Limit (§7.7.1).
requirement (e.g. "must be under 0.05 mg/kg lead"). → *"Does the customer's result meet the
specification?"* That's an Acceptance Criterion (§7.8.6).
Same-shaped numbers, completely different meaning.
It's the specification the CUSTOMER's result is judged against to issue a **statement of
conformity** (conforms / does not conform).
limit, or a product standard.
"Why decision rules and conformity statements").
Important: most agronomic parameters (e.g. the Ca/Mg ratio, C/N, available boron) have no
conformity specification. They don't "pass/fail" — they are interpreted (low / adequate / high)
via the Ranges and the Observations/Recommendations. For those, the CONFORMITY column correctly
shows "—" (Not Applicable).
These are the control limits (target ± limits, Shewhart-style) for your own control material —
the known sample you run regularly to prove the method is in statistical control.
difference between an original and its replicate.
| Acceptance Criterion (§7.8.6) | Control-Chart Limit (§7.7.1) | |
|---|---|---|
| Subject | the customer's result | the lab's control material |
| Question | "Does it meet the spec?" | "Is the process in control?" |
| When you set it | only if a real spec exists | for every method/element you monitor with QC |
| What it drives | CONFORMITY + conformity statement | Control Sample pass/fail + RPD check |
| Screen | Quality Control → Acceptance Criteria | Quality Control → Control-Chart Limits |
agronomic parameters). The limits actually applied appear beneath the icon, so the ✓/✗ explains itself.
If you're setting limits to check whether your method is working → Control-Chart Limit.If you're setting limits to state whether the customer's result meets a requirement →
Acceptance Criterion.
Don't register your control material's limits as Acceptance Criteria: then every customer's result is
wrongly judged "non-conforming" against a limit that was only ever about your own control sample.