Documentation
Tutorials
Workflows
Kebab actions
Daily Work
Dashboard & Reports
Quality Control
ISO 17025
Reagents & Storage
Logs
Contacts
Billing
Analysis Setup
Settings
Initial Setup
Help & Reference
Standards
Architecture decisions
Positioning
Outage continuity · Part 2/3
2·Introduction0:000:00

Work during an outage

This guide describes how the laboratory keeps working when it cannot reach the app — in a power cut, an internet cut, or when the app itself is unavailable. The forms pack and the labels are prepared beforehand (see Prepare for an outage), and the paper is entered into the app afterwards (see After an outage).


The basic principle

The paper filled in during the outage is the original record. The later entry in the app is a copy of it. That is why the form is filled in at the time of the work, carries a date and initials, and is kept after it has been entered (§7.5.1, §8.4.2).

The paper has to exist before the outage: in a power cut nothing can be printed.


What works in each case

CaseAppWhat the lab still has
Power cut at the labWorks (it runs on the server)No computers and no printer; often no network. Phones on mobile data work.
Internet cut at the labWorksComputers and printer work, without a connection. Phones on mobile data work.
The app is unavailableDoes not workEverything works except the app.

While the app works, a phone on mobile data reaches it as a computer would, and whatever is entered from there needs no paper. When the app itself is unavailable, the phone does not help: work is done on paper only.

Continues by hand: receiving samples, field sampling, preparation (drying, grinding, sieving, extraction), readings with battery-powered or manual equipment, environmental readings, recording instrument and reagent lot use, recording a deviation or a customer request.

Waits until service is restored: readings on instruments that need power or the lab's computers, reports, invoices and emails.


Which form for which task

TaskForm
Receiving one sampleSample Receipt (SRF) — stick the label on the sample and write the same number in the "Sample number (pre-printed label)" field
Many samples at the doorSample Reception Log — one line per sample
Readings at the benchThe Blank Worksheet of the corresponding method
Sampling in the fieldThe Field Sampling Form of the sampling plan
Temperature and humidity of storage locationsEnvironmental Readings Sheet
Drying, grinding, sievingSample Preparation Sheet
Use of an instrumentInstrument Use Log
Use of a reagent lotReagent Lot Use Log
Deviation or nonconformityDeviation / Nonconformity Note
Customer request or complaintCustomer Request or Complaint Note
Sending a sampleShipment / Chain-of-Custody Note

The reference sheets — Acceptance Criteria by Method, Instrument Calibration Validity, Authorised Staff by Method — are not filled in; they let staff judge by hand whether a result is acceptable, whether an instrument is fit for use, and who is authorised for a method.

How the paper is filled in

  • On every form, fill in the "Emergency form" line: Outage date and Page.
  • Write at the moment the work is done, with the date and time, not later from memory (§7.5.1).
  • Every form ends with "Filled in by (name, signature)" and a date.
  • A correction is made with a single line through the error, so the original value stays readable, with initials and date; never by erasing or writing over it (§7.5.2).
  • Do not invent sample numbers: use only the pre-printed labels.

Field sampling from a phone

On a phone where the Sample Collection page is already open (Daily Work → Sample Collection), entry continues without a connection: the sample and its photos are kept on the device and sent when there is a connection again. Whatever has not yet reached the server shows in the Sync Queue. Details in Sample Collection and Sync Queue. This covers field sampling only; for everything else, paper applies.


See also